1
2
*Corresponding Author E-mail: vaishududhabale1995@gmail.com
A simple and precise method was developed for the assay of Ivabradine from SR tablet formulation. The solvent system and wavelength were optimized in order to maximize the sensitivity of the proposed method; Ivabradine shows the maximum absorbance at 286 nm. The separation was achieved on RP-HPLC Isocratic system equipped with HPLC Agilent 1100 series. The mobile phase was prepared with buffer pH 5.8: Acetonitrile: Methanol in the ratio of 65: 25: 10 o-phosphoric acid used for the pH adjustment (pH-5.8). The method was validated for accuracy, precision, linearity, Specificity, stressed studies, robustness, etc. Linearity was observed in the concentration range of 70–130%and gave mean correlation coefficient 0.999. The developed RP-HPLC method was found to be accurate, precise and was successful applied to a pharmaceutical tablet formulation for qualitative estimation of Ivabradine.
Ivabradine, RP-HPLC, Force degradation, method validation