A simple precise, rapid accurate and sensitive non-aqueous potentiometric titration method was developed for quantitative determination of isoxsuprine hydrochloride from pharmaceutical dosage form. The titration was carried out using standardized 0.1 N perchloric acid. The proposed method was found to be precise with %RSD <1 (n = 6). The method showed strict linearity (r2 > 0.9999) between 10% to 50% of 0.500mg of drug substance weight. The percentage recovery of isoxsuprine hydrochloride in the optimized method was between 98.66 to 99.95%. The method is also found to be rugged when checked by different analysts and using different lots of reagents and different makes of titrators.
Isoxsuprine hydrochloride, Perchloric acid, Potassium hydrogen phthalate, Glacial acetic acid mercury acetate