Asian Journal of Research in Chemistry
  • Year: 2010
  • Volume: 3
  • Issue: 1

Development and Validation of an HPLC Method for Analysis of Gemifloxacin in Human Plasma

  • Author:
  • GVH Raju, S Ganapathy, DG Sankar, PY Naidu
  • Total Page Count: 5
  • Page Number: 192 to 196

Department of Pharmaceutical Sciences, Andhra University, Visakhapatnam-530003

*Corresponding Author E-mail: gadepalli_raj@yahoo.com

Online published on 13 March, 2013.

Abstract

A simple high-performance liquid chromatographic method for the determination of Gemifloxacin in human plasma has been developed. Separation was achieved by Reverse phase chromatography on a Purospher RP18e(150 x 4.6 mm, 5 μm) column with mobile phase composition of pH 3.0 buffer: acetonitrile: Methanol in the ratio 75:17:8 v/v with ultra violet detection at 273nm. Lower limit of quantitation was 50 ng/ml. Maximum between-run precision was 7.44%. Mean extraction recovery was found to be 92.53 to 109.87%. Stability study showed that after three freezethaw cycles the loss of three quality control samples were less than 10%. Samples were stable at room temperature for 48h and at −20° for 2months. Before injecting onto HPLC system, the processed samples were stable for at least 6h. The method was used to perform bioequivalence study in human.

Keywords

RP HPLC, Validation, Atenolol