Government College of Pharmacy, Vedant Road, Osmanpura, Aurangabad-431005
*Corresponding Author E-mail: prabhakarpanzade@gmail.com
Online published on 13 March, 2013.
A simple, accurate, cost effective and reproducible spectrophotometric method has been developed for the estimation of Granisetron hydrochloride in bulk and pharmaceutical dosage form. The developed UV spectrophotometric method is based on measurement of absorption at maximum wavelength 302.80 nm. The developed method was validated with respect to linearity, precision, accuracy and specificity, according to the ICH guidelines. Beer's law was obeyed in the concentration range of 5–30 μg/ml having line equation y = 0.03315x + 0.01533 with correlation coefficient of 0.99981. The percentage recovery of Granisetron hydrochloride ranged from (101.6298±1.0547) in pharmaceutical dosage form. Results of the analysis were validated statistically and by recovery study.
UV Spectroscopy, percentage recovery, Granisetron hydrochloride, ICH guidelines