Asian Journal of Research in Chemistry
  • Year: 2011
  • Volume: 4
  • Issue: 9

A Simple and Sensitive RP-HPLC Method for Estimation of Trandolapril in Bulk and Tablet Dosage Forms

  • Author:
  • V. Bhaskara Raju1, A. Lakshmana Rao2,
  • Total Page Count: 3
  • Page Number: 1425 to 1427

1Sri Vasavi Institute of Pharmaceutical Sciences, Tadepalligudem-534 101, Andhra Pradesh, India

2V.V. Institute of Pharmaceutical Sciences, Gudlavalleru-521 356, Andhra Pradesh, India

*Corresponding Author E-mail: dralrao@gmail.com

Online published on 6 March, 2013.

Abstract

An accurate and precise HPLC method was developed for the determination of trandolapril. Separation of the drug was achieved on a reverse phase C18 column using a mobile phase consisting of phosphate buffer and acetonitrile in the ratio of 35:65 v/v. The flow rate was 0.8 ml/min and the detection wavelength was 210 nm. The linearity was observed in the range of 20–60 μg/ml with a correlation coefficient of 0.999. The proposed method was validated for its linearity, accuracy, precision and robustness. This method can be employed for routine quality control analysis of trandolapril in tablet dosage forms.

Keywords

Trandolapril, Estimation, RP-HPLC, Validation, Tablets