Parul Institute of Pharmacy, Waghodia, Limda, Gujarat, India
*Corresponding Author E-mail: ujjaval18@gmail.com
Online published on 21 September, 2013.
An Absoption Correction UV spectrophotometric and a reversed Phase high-performance liquid chromatographic method for the determination of Ambroxol and Olopatadine Hydrochloride in Synthetic mixture were developed. The Absorption Correction UV spectrophotometric method was found to be accurate with 99.75±0.04 recovery for ambroxol and recovery for olopatadine 99.89±0.02 respectively and method is also precise with a coefficient of variation for olopatadine and ambroxol was 0.996 and 0.885 respectively. For reverse phase HPLC, A reversed-phase C18 column with methanol: ammonium acetate buffer (80:20, v/v, pH 5.5) mobile phase was used and the detector wavelength was set at 244 nm. Linearity range for ambroxol and Olopatadine Hcl was found to be 78–150 μg/ml and 13–25 μg/ml respectively.
Ambroxol, Olopatadine Hydrochloride, Absoption Correction UV spectrophotometry, Reversed phase HPLC