In vitro Biomechanical Testing of Polyurethane External Skeletal Fixator as a Novel Polymer in Fracture Treatment

  • Author:
  • Ünal Yavuz1, Kerem Yener1,*, Mehmet Salih Karadağ1, Kübra Dikmen İLGİNOĞLU11, Mehmet Sıdık Hurma1, Ali Hayat1, Mustafa Özen2
  • Total Page Count: 8
  • Page Number: 1800 to 1807

1Department of Surgery, Faculty of Veterinary Medicine, University of Harran, Sanliurfa, Türkiye

2Department of Mechanical Engineering, Faculty of Engineering, University of Harran, Şanlıurfa, Türkiye

*Corresponding Author: Kerem Yener, Department of Surgery, Faculty of Veterinary Medicine, University of Harran, Şanlıurfa, Türkiye. Email: keremyener@harran.edu.tr

Abstract

To determine the temperature, time and biomechanical properties of free-form polyurethane external skeletal fixation (PUESF) and compare them with acrylic external skeletal fixation (ACESF).

Three different configurations (uniplanar, multiplanar and circular) (n=4 for each configuration) were subjected to compression and bending (n=24 PUESF, n=24 ACESF) and only the uniplanar configuration was subjected to pullout testing (n=4 PUESF, n=4 ACESF). The temperatures of the connecting bar and K-wire were recorded when creating the multiplanar and circular configurations. All the configurations were subjected to axial and bending tests and the uniplanar configuration was subjected to pullout tests.

Regarding the connecting bar temperature and the temperature delivered to the K-wires, the PUESF group reached maximum temperatures of 75.7°C and 38.1°C at 3 minutes and the ACESF group reached maximum temperatures of 87.6°C and 60°C at 9 minutes, respectively (p<0.001). Under compression force, the PUESF group showed similar strength to the ACESF group in the uniplanar and circular configurations (p>0.05), while it showed less strength than the ACESF group in the multiplanar configuration (p<0.05). The PUESF group showed similar strength to the ACESF group in all three configurations under the bending force (p>0.05). The ACESF group showed approximately 2 times as much pullout strength as that of the PUESF group (p<0.05). Polyurethane showed acceptable mechanical performance compared to acrylic in several test configurations. Polyurethane can be used in clinical trials for free-form external skeletal fixations (ESFs).

Keywords

Acrylic, Biomechanical testing, External skeletal fixation, Polyurethane