1Flat B-203, Block-B, United Avenue Apartments, South End, 7-1-29, Ameerpet, Hyderabad-500016
21-2-98, Kakatiyanagar, Habsiguda, Hyderabad-500007
*Corresponding Author E-mail: ksprao1939@yahoo.co.in
Online published on 7 August, 2012.
The paper proposes a simple, accurate, precise, rapid, and non-destructive wavelength-dispersive x-ray fluorescence spectrometric technique for determining vanadium, chromium, cobalt, nickel, copper, zinc, barium, and lead in granitic rocks. The technique uses a vacuum x-ray spectrograph, 60 kV- 40 mA- 2 kW x-ray generator, tungsten x- ray tube, LiF 200 analysing crystal, coarse collimator, vacuum path, gas-flow proportional counter, short counting times, and the international rock standards, G-1, G-2, GSP-1, DTS-1, PCC-1, BCR-1, W-1, and AGV-1.
The accuracy of the proposed technique is excellent (within 1% for chromium, copper, zinc, barium, and lead; within 2% for vanadium and nickel) and very good (within 4% for cobalt). The precision is also excellent (within 1% for copper, zinc, barium, and lead; within 2% for vanadium, chromium, and nickel) and very good (within 5% for cobalt). The lower limits of detection for the eight elements are: 1 ppm for chromium, cobalt, nickel, copper, and zinc; 2 ppm for vanadium and lead; and 5 ppm for barium. The time taken for determining vanadium, chromium, cobalt, nickel, copper, zinc, barium, and lead in a batch of twentyfour samples of granitic rocks, for a replication of four analyses per sample, by one operator, using a manual x-ray fluorescence spectrometer, is only twentyfour hours.
X-ray fluorescence spectrometry, WDXRFS, Granitic rocks, Vanadium, Chromium, Cobalt, Nickel, Copper, Zinc, Barium, Lead