Ultrastructural Detection of Stainless Steel and Titanium Metal Particles in Soft Tissues - Used in Internal Fixation for Maxillofacial Fractures

  • Author:
  • A Chris Deepak1, R Dilip Kumar2, HP Raghuveer3, Jagadish Sarvadnya4
  • Total Page Count: 3
  • Page Number: 47 to 49

1Resident, Dept. of Oral Oncology and Reconstructive Surgery, Kidwai Memorial Institute of Oncology, Bangalore, India

2Professor, Dept. Oral and Maxillofacial Surgery, Dayananda Sagar College of Dental Sciences, Bangalore, India

3Principal and HOD, Dept. of Oral and Maxillofacial Surgery, Dayananda Sagar College of Dental Sciences, Bangalore, India

4Fellow, Dept. of Oral Oncology and Reconstructive Surgery, Kidwai Memorial Institute of Oncology, Bangalore, India

Online published on 29 August, 2013.

Abstract

The aim of the study was to analyze the tissue response to stainless steel and titanium mini plates used in maxillofacial trauma.

30 patients treated for various mandibular fractures were considered: 12 patients treated with stainless steel plate and 18 treated with titanium plates, time gap varying from 10 to 79 months.

Transmission electron microscopic results reported leaching of metallic particles, which were observed in 20 patients, of whom 8 were plated with titanium plates and 12 with stainless steel plates. Leaching of metallic particles reported in these 20 patients had implantation period of more than 15 months. Mixture of chronic inflammatory cells was found around the implant on compound microscopy.

Leaching of metallic particles into the surrounding tissues more commonly occurs with stainless steel plates and also in patients in whom plates were retained for a prolonged period of time of more than 15 months. Plates should be removed once their purpose is fulfilled.

Keywords

Transmission electron microscope, Stainless steel plates, Titanium plates