Research Journal of Pharmacy and Technology
SCOPUS
  • Year: 2013
  • Volume: 6
  • Issue: 7

Development and Validation of RP-HPLC Method for Estimation of Levosalbutamol Sulphate and Ipratropium Bromide in Pressured Meter Dose Inhaler Form

  • Author:
  • Nandini R. Pai, Swapnali Suhas Patil
  • Total Page Count: 6
  • Page Number: 774 to 779

Department of Chemistry, D.G. Ruparel College, Mahim, Mumbai-400016, Maharashtra, India

*Corresponding Author E-mail: nandini_pai@hotmail.com

**swapnali22@gmail.com

Online published on 31 October, 2013.

Abstract

Simple and sensitive RP-HPLC methods are described for the determination of Levosalbutamol sulphate and Ipratropium bromide combination. The high-performance liquid chromatographic method was developed for the determination of Levosalbutamol sulphate and Ipratropium bromide Pressurized meter dose inhaler (LI-MDI) using acetonitrile: buffer [30: 70] as the mobile phase and measuring the response at λmax 210 nm. The analysis was performed on a Chromatopak C18 (150 X 4.6 mm), 5-μm column. The method was found to be accurate, with linearity ranging from (80.0% to 120%) for HPLC with a correlation coefficient (r2) 0.999. Precision mean % assay was found with in range.

Keywords

RP-HPLC, Levosalbutamol sulphate, Ipratropium bromide, Pressurised Meter Dose Inhaler